Red-lion CUB5RT Bedienungsanleitung

Stöbern Sie online oder laden Sie Bedienungsanleitung nach Computerzubehör Red-lion CUB5RT herunter. Red Lion CUB5RT User Manual Benutzerhandbuch

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 12
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 0
1
MODEL CUB5RT - MINIATURE ELECTRONIC 5-DIGIT RTD METER
GENERAL DESCRIPTION
The CUB5 provides the user the ultimate in flexibility, from its complete user
programming to the optional setpoint control and communication capability.
The CUB5RT accepts an RTD input and provides a temperature display in
Celcius or Farenheit. The meter also features minimum and maximum display
capture, display offset, °F or °C indicator, and programmable user input. The
display can be toggled either manually or automatically between the selected
displays.
The CUB5 display has 0.48" (12.2 mm) high digits. The LCD is available in
two versions, reflective and red/green backlight. The backlight version is user
selectable for the desired color and also has variable display intensity.
The capability of the CUB5 can be easily expanded with the addition of
option cards. Setpoint capability is field installable with the addition of the
setpoint output cards. Serial communications capability for RS232 or RS485 is
added with a serial option card.
The CUB5 can be powered from an optional Red Lion Micro-Line/Sensor
Power Supply (MLPS), which attaches directly to the back of a CUB5. The
MLPS is powered from 85 to 250 VAC and provides up to 400 mA to drive the
unit and sensors.
SAFETY SUMMARY
All safety related regulations, local codes and instructions that appear in this
literature or on equipment must be observed to ensure personal safety and to
prevent damage to either the instrument or equipment connected to it. If
equipment is used in a manner not specified by the manufacturer, the protection
provided by the equipment may be impaired.
Do not use this meter to directly command motors, valves, or other actuators
not equipped with safeguards. To do so can be potentially harmful to persons or
equipment in the event of a fault to the meter.
Bulletin No. CUB5RT-H
Drawing No. LP0644
Released 02/13
Tel +1 (717) 767-6511
Fax +1 (717) 764-0839
www.redlion.net
CAUTION: Risk of Danger.
Read complete instructions prior to
installationand operation of the unit.
CAUTION: Risk of electric shock.
DIMENSIONS In inches (mm)
.13 (3.3)
(43.4)
1.71
2.95 (74.9)
SEL
.15 (3.8)
1.54 (39.1)
RST
(68 )
+.025
-.000
+8
-0
1.29 (32.8)
2.68
-.000
1.30
(33 )
-0
+6
+.024
Note: Recommended minimum clearance (behind the panel) for mounting clip installation is 2.15" (54.6) H x 3.00" (76.2) W.
C
C
US LISTED
US LISTED
U
L
R
51EB
IND. CONT. EQ.
RTD INPUTS
RTD types Pt385, Pt392, Ni672, Cu427
PROGRAMMABLE TEMPERATURE OFFSET
SELECTABLE °F or °C WITH 1 or 0.1 DEGREE RESOLUTION
°F OR °C DISPLAY ANNUNCIATORS
MINIMUM AND MAXIMUM DISPLAY CAPTURE
LCD, REFLECTIVE OR GREEN/RED LED BACKLIGHTING
0.48" (12.2 mm) HIGH DIGITS
OPTIONAL SETPOINT OUTPUT CARD
OPTIONAL SERIAL COMMUNICATION CARD (RS232 or RS485)
OPTIONAL USB PROGRAMMING CARD
OPERATES FROM 9 TO 28 VDC POWER SOURCE
FRONT PANEL OR CRIMSON PROGRAMMABLE
DISPLAY COLOR CHANGE CAPABILITY AT SETPOINT OUTPUT
NEMA 4X/IP65 SEALED FRONT BEZEL
Seitenansicht 0
1 2 3 4 5 6 ... 11 12

Inhaltsverzeichnis

Seite 1 - DIMENSIONS In inches (mm)

1MODEL CUB5RT - MINIATURE ELECTRONIC 5-DIGIT RTD METERGENERAL DESCRIPTIONThe CUB5 provides the user the ultimate in flexibility, from its complete use

Seite 2 - ORDERING INFORMATION

10SETPOINT VALUEEnter the desired setpoint value. The decimal point position for the setpoint and hysteresis values follow the selection set in Module

Seite 3 - OPTIONAL PLUG-IN CARDS

116.5 MODULE 5 - SERIAL SETUP PARAMETERS (5-SEr)Enter the probe burn-out action. In the event of a temperature probe failure (open or short), the out

Seite 4 - 3.0 INSTALLING PLUG-IN CARDS

Setpoint ActionSetpoint SelectSetpoint ValueOutput Reset ActionHysteresis ValueOff Time DelayOn Time DelayOutput Reset With Display ResetChange Displ

Seite 5 - (30 V MAX.)

1. DISPLAY: 5 digit LCD 0.48" (12.2 mm) high digitsCUB5RTR0: Reflective LCD with full viewing angleCUB5RTB0: Transmissive LCD with selectable red

Seite 6 - 6.0 PROGRAMMING THE METER

3ADDING OPTION CARDSThe CUB5 meters can be fitted with optional output cards and/or serial communications cards. The details for the plug-in cards can

Seite 7 - This can be

WIRING OVERVIEWElectrical connections are made via screw-clamp terminals located on the back of the meter. All conductors should conform to the meter’

Seite 8 - LO-En LO-t FCS

54.1 POWER WIRINGCOMMUSR+9-28 VDCEXCINP+PWR COMMON+-DC Power+9 to +28 VDC: +VDCPower Common: -VDC4.2 USER INPUT WIRINGCOMMUSR+9-28 VDCEXCINP+PWR COM

Seite 9 - ARAMETERS (3-dSP)

65.0 REVIEWING THE FRONT BUTTONS AND DISPLAYSEL RSTPROGRAMMING MODE ENTRY (SEL BUTTON)It is recommended that all programming changes be made off line,

Seite 10 - Stb-n brn-n

Temperature ScaleDisplay Decimal PointFilter SettingUser Input AssignmentRTD TypeSELtYPESCALEdECPt OFSEt FILtrbANdU-ASN1-INPProUser Input FunctionDisp

Seite 11 - PAr Addr Abbr

86.2 MODULE 2 - SECONDARY FUNCTION PARAMETERS (2-SEC)SELMax Capture Delay TimeMax Display EnableMin Display EnableAccess Code For Service OperationsM

Seite 12 - Press and hold SEL button

9The Security Code determines the programming mode and the accessibility of programming parameters. This code can be used along with the Program Mode

Kommentare zu diesen Handbüchern

Keine Kommentare